Morphological changes in the iliac artery walls of pigs after implantation of spiral devices from β-zirconium alloy

  • V. P. Zakharova National M. M. Amosov Institute of Cardiovascular Surgery National Academy of Medical Sciences of Ukraine
  • Yu. V. Panichkin National M. M. Amosov Institute of Cardiovascular Surgery National Academy of Medical Sciences of Ukraine
  • I. O. Skiba G. V. Kurdyumov Institute of Metal Physics of the National Academy of Sciences of Ukraine (Kiev)
  • M. L. Zhuravleva National M. M. Amosov Institute of Cardiovascular Surgery National Academy of Medical Sciences of Ukraine
  • V. M. Beshlyaga National M. M. Amosov Institute of Cardiovascular Surgery National Academy of Medical Sciences of Ukraine
  • E. V. Beshlyaga National M. M. Amosov Institute of Cardiovascular Surgery National Academy of Medical Sciences of Ukraine
  • Yu. A. Ruzhin National M. M. Amosov Institute of Cardiovascular Surgery National Academy of Medical Sciences of Ukraine
Keywords: patent ductus arteriosus, a heart occluder, the biological model, β-zirconium alloy

Abstract

The results of histological studies of the iliac arteries of pigs wall after the implantation of the spiral device – occluder, made on the basis of low modulus β-zirconium alloy. It is shown that in 2.5 months after the introduction into the lumen of the artery of the occluder, a capsule is formed around of neointimal fibrosis, which overlaps occluder more than 2/3 of the lumen of the vessel, indicating a good device occlusive effect.

References

1. Porstmann W., Wierny L., Warnke H. Catheter closure of patent ductus arteriosus: 62 cases treated without thoracotomy // Radiol Clin North Am. – 1971. – Vol. 9. – Р. 203–218.

2. Celiker A., Aypar E., Karagoz T., Dilber E., Ceviz N. Transcatheter closure of patent ductus arteriosus with Nit-Occlud coils // Catheter Cardiovasc Interv. – 2005. – Vol. 65. – Р. 569–76.

3. Gruenstein D. H., Bass J. L. Experimental evaluation of a new articulated Amplatzer ductal occlude device without fabric // Catheter Cardiovascular Interventional. – 2009. – Vol. 74. – Р. 482–487.

4. Ryhanen J., Niemi E., Serlo W., Niemela E., Sandvik P., Pernu H., et al. Biocompatibility of nickel-titanium shape memory metal and its corrosion behavior in human cell cultures // J Biomed Mater Res. – 1997. – Vol. 35. – Р. 451–457.

5. Niinomi M. Mechanical biocompatibilities of titanium alloys for biomedical applications // Journal of the Mechanical Behavior of Biomedical Materials. – 2000. – Vol. 1. – P. 30–42.

6. Trepanier C., Venugopalan R., Messer R. Z. J., Pelton A. R. Effect of Passivation treatments on nickel release from nitinol // Society for Biomateria. – 6th World Biomaterials Congress, Transactions. – 2000. – Р. 1043.

7. Burian M., Neumann T., Weber M., Brandt R., Geisslinger G., Mitrovic V., et al. Nickel release, a possible indicator for the duration of antiplatelet treatment, from a nickel cardiac device in vivo: a study in patients with atrial septal defects implanted with an Amplatzer occlude // Int J. Clin.Pharmacol Ther. – 2006. – Vol. 44. – Р. 107–112.

8. Sigler M., Jux C. Biocompatibility of septal defect closure devices // Heart. – 2007. – Vol. 9. – Р. 444–449.

9. Скиба И. А., Карасевская О. П., Мордюк Б. Н., Марковский П. Е., Шиванюк В. Н. Влияние инициируемого деформацией ?>?превращения на механическое поведение β-сплавов титана и циркония // Металлофизика и новейшие технологии. – Т. 31. – № 1. – 2009. – С. 1573–1586.

10. Кулеш Д. В., Скиба И. О., Карасевская О. П. и др. Микроструктура, механические свойства и биосовместимость нового Zr-Ti-Nb сплава // Пластична та реконструктивна хірургія. – 2011. – № 2 (XVII). – С. 44–50.

11. Паничкин Ю. В., Скиба И. А., Захарова В. П. и др. Особенности методики проведения доклинического эксперимента по имплантации окклюдера из β-циркониевого сплава на свиньях как биологической модели // Серце і судини. – 2015. – № 4 (52). – С. 25–30.
Published
2016-09-12
How to Cite
1.
Zakharova VP, Panichkin YV, Skiba IO, Zhuravleva ML, Beshlyaga VM, Beshlyaga EV, Ruzhin YA. Morphological changes in the iliac artery walls of pigs after implantation of spiral devices from β-zirconium alloy. ujcvs [Internet]. 2016Sep.12 [cited 2025Feb.5];(2 (25):105-7. Available from: https://cvs.org.ua/index.php/ujcvs/article/view/248